《分流不分流进样口介绍.ppt》由会员分享,可在线阅读,更多相关《分流不分流进样口介绍.ppt(14页珍藏版)》请在三一文库上搜索。
1、Slide #1,6850/6890 Split Animation Split/Splitless Inlet - Component Identification,Septum,Liner,Liner O-Ring,Gold Seal,Capillary Column,5/4/00,Slide #2,6850/90 Split/Splitless Inlet - Component Identification,Septum,SEPTUM PURGE VENT,Inlet Split Port,Slide #3,6850/90 Split Animation - Default Condi
2、tions,FLOW SENSOR,PROPORTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,SPLIT VENT,GOLD SEAL,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,0 ML/MIN,SPLIT VENT TRAP,0 PSI,Slide #4,6850/90 Split Animation - Total Flow Setting,FLOW SENSOR,PROPORTIONAL VALVE 1
3、,PRESSURE SENSOR,SEPTUM PURGE REGULATOR,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,SPLIT VENT 205 ML/MIN,GOLD SEAL,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,PRIMARY FLOW PATH,REDUCED FLOW PATH,205 ML/MIN,SPLIT VENT TRAP,0 PSI,Slide #5,6850/90 Split Animation - Inlet Pressure Setting,FLOW SENSOR,PROPO
4、RTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FIXED 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,GOLD SEAL,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,PRIMARY FLOW PATH,REDUCED FLOW PATH,205 ML/MIN,SPLIT VENT TRAP,10 PSI,SPLIT VENT 200 ML/MIN,COLUMN FLOW 2 ML/MIN,PRESSURIZED GAS,BACK PRE
5、SSURE CONTROL LOOP,Slide #6,6850/90 Split Ready for Injection Animation,FLOW SENSOR,PROPORTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FIXED 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,205 ML/MIN,SPLIT VENT TRAP,10 PSI,SPLIT VENT 200 ML/MIN,CO
6、LUMN FLOW 2 ML/MIN,BACK PRESSURE CONTROL LOOP,Split Ratio - 100:1,Mode: Split Temp 250 250 Pressure 10.0 10.0 Split Ratio 100 Split Flow 200 Tot Flow 205.0 205.0 Gas Saver OFF,Slide #7,6850/90 Split Insert Syringe Animation,FLOW SENSOR,PROPORTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FIXE
7、D 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,GOLD SEAL,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,205 ML/MIN,SPLIT VENT TRAP,10 PSI,SPLIT VENT 200 ML/MIN,COLUMN FLOW 2 ML/MIN,BACK PRESSURE CONTROL LOOP,Split Ratio - 100:1,Slide #8,6850/90 Split Inject Sample Animation,FLOW SENSOR,PROPORTIONAL
8、VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FIXED 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,GOLD SEAL,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,205 ML/MIN,SPLIT VENT TRAP,10 PSI,SPLIT VENT 200 ML/MIN,COLUMN FLOW 2 ML/MIN,BACK PRESSURE CONTROL LOOP,Split Ratio - 100:1,Slide #9,6850/90 Spli
9、t Sample Vaporizes Animation,FLOW SENSOR,PROPORTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FIXED 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,GOLD SEAL,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,205 ML/MIN,SPLIT VENT TRAP,10 PSI,SPLIT VENT 200 ML/MIN,COLUMN FLOW 2 ML/MIN,BACK PRESSURE
10、CONTROL LOOP,Split Ratio - 100:1,Slide #10,6850/90 Split Sample is Split - Animation Good Liner Design?,FLOW SENSOR,PROPORTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FIXED 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,205 ML/MIN,SPLIT VENT TRAP
11、,10 PSI,SPLIT VENT 200 ML/MIN,COLUMN FLOW 2 ML/MIN,BACK PRESSURE CONTROL LOOP,Split Ratio - 100:1,Agilent Split Liner - 5183-4647,Slide #11,6850/90 Split Split Continues Animation,FLOW SENSOR,PROPORTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FIXED 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTRO
12、L LOOP,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,205 ML/MIN,SPLIT VENT TRAP,10 PSI,SPLIT VENT 200 ML/MIN,COLUMN FLOW 2 ML/MIN,BACK PRESSURE CONTROL LOOP,Split Ratio - 100:1,Slide #12,6850/90 Split Split is Complete Animation,FLOW SENSOR,PROPORTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FI
13、XED 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,205 ML/MIN,SPLIT VENT TRAP,10 PSI,SPLIT VENT 200 ML/MIN,COLUMN FLOW 2 ML/MIN,BACK PRESSURE CONTROL LOOP,Split Ratio - 100:1,Slide #13,6850/90 Split Sample is separated Animation and Detected,FLOW SENSOR,P
14、ROPORTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FIXED 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,205 ML/MIN,SPLIT VENT TRAP,10 PSI,SPLIT VENT 200 ML/MIN,COLUMN FLOW 2 ML/MIN,BACK PRESSURE CONTROL LOOP,Split Ratio - 100:1,Slide #14,6850/90 S
15、plit Gas Saver - Set to 55 ml/min Animation After the Injection is complete,FLOW SENSOR,PROPORTIONAL VALVE 1,PRESSURE SENSOR,SEPTUM PURGE REGULATOR FIXED 3ML/MIN,PURGE VALVE OPEN,TOTAL FLOW CONTROL LOOP,INPUT PRESSURE,PROPORTIONAL VALVE 2,80 PSI,55 ML/MIN,SPLIT VENT TRAP,10 PSI,SPLIT VENT 50 ML/MIN,COLUMN FLOW 2 ML/MIN,BACK PRESSURE CONTROL LOOP,Split Ratio - 100:1,Mode: Split Temp 250 250 Pressure 10.0 10.0 Split Ratio 100 Split Flow 200 Tot Flow 205.0 205.0 Gas Saver ON Saver Flow 50.0 Saver Time 0.5,
链接地址:https://www.31doc.com/p-3112205.html