BS EN 14626-2005 环境空气质量.用非色散红外线光谱法测量一氧化碳浓度的标准方法.pdf
《BS EN 14626-2005 环境空气质量.用非色散红外线光谱法测量一氧化碳浓度的标准方法.pdf》由会员分享,可在线阅读,更多相关《BS EN 14626-2005 环境空气质量.用非色散红外线光谱法测量一氧化碳浓度的标准方法.pdf(90页珍藏版)》请在三一文库上搜索。
1、BRITISH STANDARD BS EN 14626:2005 Ambient air quality Standard method for the measurement of the concentration of carbon monoxide by nondispersive infrared spectroscopy The European Standard EN 14626:2005 has the status of a British Standard ICS 13.040.20 BS EN 14626:2005 This British Standard, was
2、published under the authority of the Standards Policy and Strategy Committee on 31 March 2005 BSI 31 March 2005 ISBN 0 580 45718 4 National foreword This British Standard is the official English language version of EN 14626:2005. The UK participation in its preparation was entrusted by Technical Com
3、mittee EH/2, Air quality, to Subcommittee EH/2/3, Ambient atmospheres, which has the responsibility to: A list of organizations represented on this subcommittee can be obtained on request to its secretary. Cross-references The British Standards which implement international or European publications
4、referred to in this document may be found in the BSI Catalogue under the section entitled “International Standards Correspondence Index”, or by using the “Search” facility of the BSI Electronic Catalogue or of British Standards Online. This publication does not purport to include all the necessary p
5、rovisions of a contract. Users are responsible for its correct application. Compliance with a British Standard does not of itself confer immunity from legal obligations. aid enquirers to understand the text; present to the responsible international/European committee any enquiries on the interpretat
6、ion, or proposals for change, and keep the UK interests informed; monitor related international and European developments and promulgate them in the UK. Summary of pages This document comprises a front cover, an inside front cover, the EN title page, pages 2 to 86, an inside back cover and a back co
7、ver. The BSI copyright notice displayed in this document indicates when the document was last issued. Amendments issued since publication Amd. No. DateComments W - - - - - 7 7 -7 - 7 - 90 % a performance criterion is set both at zero and span level b a H2O-concentration of 19 mmol/mol is about 80 %
8、RH at 293 K and 101,3 kPa c span level is 70 % to 80 % of the certification range d if relevant NOTE mol/mol = ppm; nmol/mol = ppb 8.3 Design changes The manufacturer shall report any (design) change of the previous type approved monitor to the compenent authority. The underlying idea of the (design
9、) change, as well as the expected consequences with respect to the performance characteristics as stated in Table 1, shall be reported by the manufacturer. The responsible authority shall decide whether or not new tests are necessary. Excluded from reporting are changes in: material of the instrumen
10、t frame (housing); electronic connectors; bolts, nuts, cable binders and other fixing materials. 8.4 Procedures for determination of the performance characteristics during the laboratory test 8.4.1 General A designated body shall perform the determination of the performance characteristics in the la
11、boratory as a part of the type approval test. The quality of the materials and equipment used in the described test procedures shall be in accordance with the requirements given in this document. The tests shall be performed on at least two analysers in the laboratory test. 8.4.2 Test conditions 8.4
12、.2.1 General Before operating the analyser, the operating instructions of the manufacturer shall be followed particularly with regard to the set-up of equipment and the quality and quantity of the consumable products necessary. EN 14626:2005 (E) 20 The analyser should be allowed to warm up during th
13、e time specified by the manufacturer before undertaking any tests. If the warm-up time is not specified, a minimum of 4 h is recommended. When applying test gases to the analyser, the test gas system shall be operated sufficiently long before starting the tests in order to stabilise the concentratio
14、ns applied to the analyser. Most analyser systems are able to give an output signal as a moving average over an adjustable period of time and some systems automatically change this integration time as a function of the frequency of the fluctuations in concentration of the detected pollutant. These o
15、ptions are typically used in order to smooth output data. It needs to be demonstrated that the set value for the averaging time or the use of an active filter will not influence the result of the averaging test and response time test. During laboratory and field tests for the type approval the setti
16、ngs of the monitor shall be as the manufacturer requires. All settings shall be noted down in the test report. 8.4.2.2 Parameters During the test for each individual performance characteristic, the values of the following parameters shall be stable within the specified range given in Table 2. Table
17、2 Set points and stability of test parameters Parameter Set points Stability Sample gas pressure Manufacturers specification (except for the sample gas pressure test in, see 8.4.7) 0,2 kPa Sample gas temperature 20 C to 23 C (except for the sample gas temperature test, see 8.4.8) 2 C Surrounding air
18、 temperature 20 C to 23 C (except for the temperature of the surrounding air test, see 8.4.9) 2 C Electrical voltage a At nominal line voltage and within manufacturers specifications (except for the voltage dependence test, see 8.4.10) 1 % Sample flow to the analyser Manufacturers specification 1 %
19、a For an analyser operating on direct current the type approval test of voltage variation shall be carried out over the range of 10 % of the nominal voltage. 8.4.2.3 Test gases For the determination of the various performance characteristics, test gases (synthetic) air containing a certain CO concen
20、tration) traceable to national standards shall be used, unless otherwise stated in this document. Various methods for the generation of test gases are given in Table 3. Table 3 Methods for preparation of test gases Method Description Standard to be used Commercial cylinder Gas cylinder (CO gas mixtu
21、re in (synthetic) air) ISO 6143 Static dilution Preparation by means of injecting known amounts of CO in a ISO 6144 EN 14626:2005 (E) 21 Method Description Standard to be used known volume Dynamic dilution Dynamic blending of cylinder gas with (synthetic) air, zero air ISO 6145 The uncertainties in
22、zero and span gases used for the laboratory and field tests shall be proven to be insignificant. Possible contamination of zero and span gas shall not significantly influence the results of the laboratory tests. Therefore the span gases and zero gas shall meet the following specifications: Maximum p
23、ermitted uncertainties in the concentration of gases used for laboratory tests: 3 %. Specification of the purity of test gases and zero gas (expressed as absolute value) is given in Table 4. Table 4a Specification for purity of span gas Pollutant Concentration CO2 6,0 % of the measured value EN 1462
24、6:2005 (E) 36 Calibration, checks and maintenance Clause Frequency Action criteria e on the result of test Testing sample manifold - influence of pressure drop induced by the manifold pump - sample collection efficiency 9.6.4.1 9.6.4.2 At least every three years influence 1 % of measured value influ