1、Association for the Advancement of Medical Instrumentation ANSI/AAMI/ISO 22442-1:2007 Medical devices utilizing animal tissues and their derivativesPart 1: Application of risk management American National Standard Objectives and uses of AAMI standards and recommended practices It is most important t
2、hat the objectives and potential uses of an AAMI product standard or recommended practice are clearly understood. The objectives of AAMIs technical development program derive from AAMIs overall mission: the advancement of medical instrumentation. Essential to such advancement are (1) a continued inc
3、rease in the safe and effective application of current technologies to patient care, and (2) the encouragement of new technologies. It is AAMIs view that standards and recommended practices can contribute significantly to the advancement of medical instrumentation, provided that they are drafted wit
4、h attention to these objectives and provided that arbitrary and restrictive uses are avoided. A voluntary standard for a medical device recommends to the manufacturer the information that should be provided with or on the product, basic safety and performance criteria that should be considered in qu
5、alifying the device for clinical use, and the measurement techniques that can be used to determine whether the device conforms with the safety and performance criteria and/or to compare the performance characteristics of different products. Some standards emphasize the information that should be pro
6、vided with the device, including performance characteristics, instructions for use, warnings and precautions, and other data considered important in ensuring the safe and effective use of the device in the clinical environment. Recommending the disclosure of performance characteristics often necessi
7、tates the development of specialized test methods to facilitate uniformity in reporting; reaching consensus on these tests can represent a considerable part of committee work. When a drafting committee determines that clinical concerns warrant the establishment of minimum safety and performance crit
8、eria, referee tests must be provided and the reasons for establishing the criteria must be documented in the rationale. A recommended practice provides guidelines for the use, care, and/or processing of a medical device or system. A recommended practice does not address device performance per se, bu
9、t rather procedures and practices that will help ensure that a device is used safely and effectively and that its performance will be maintained. Although a device standard is primarily directed to the manufacturer, it may also be of value to the potential purchaser or user of the device as a frame
10、of reference for device evaluation. Similarly, even though a recommended practice is usually oriented towards healthcare professionals, it may be useful to the manufacturer in better understanding the environment in which a medical device will be used. Also, some recommended practices, while not add
11、ressing device performance criteria, provide guidelines to industrial personnel on such subjects as sterilization processing, methods of collecting data to establish safety and efficacy, human engineering, and other processing or evaluation techniques; such guidelines may be useful to health care pr
12、ofessionals in understanding industrial practices. In determining whether an AAMI standard or recommended practice is relevant to the specific needs of a potential user of the document, several important concepts must be recognized: All AAMI standards and recommended practices are voluntary (unless,
13、 of course, they are adopted by government regulatory or procurement authorities). The application of a standard or recommended practice is solely within the discretion and professional judgment of the user of the document. Each AAMI standard or recommended practice reflects the collective expertise
14、 of a committee of health care professionals and industrial representatives, whose work has been reviewed nationally (and sometimes internationally). As such, the consensus recommendations embodied in a standard or recommended practice are intended to respond to clinical needs and, ultimately, to he
15、lp ensure patient safety. A standard or recommended practice is limited, however, in the sense that it responds generally to perceived risks and conditions that may not always be relevant to specific situations. A standard or recommended practice is an important reference in responsible decision-mak
16、ing, but it should never replace responsible decision-making. Despite periodic review and revision (at least once every five years), a standard or recommended practice is necessarily a static document applied to a dynamic technology. Therefore, a standards user must carefully review the reasons why
17、the document was initially developed and the specific rationale for each of its provisions. This review will reveal whether the document remains relevant to the specific needs of the user. Particular care should be taken in applying a product standard to existing devices and equipment, and in applyi
18、ng a recommended practice to current procedures and practices. While observed or potential risks with existing equipment typically form the basis for the safety and performance criteria defined in a standard, professional judgment must be used in applying these criteria to existing equipment. No sin
19、gle source of information will serve to identify a particular product as “unsafe“. A voluntary standard can be used as one resource, but the ultimate decision as to product safety and efficacy must take into account the specifics of its utilization and, of course, cost-benefit considerations. Simila
20、rly, a recommended practice should be analyzed in the context of the specific needs and resources of the individual institution or firm. Again, the rationale accompanying each AAMI standard and recommended practice is an excellent guide to the reasoning and data underlying its provision. In summary,
21、 a standard or recommended practice is truly useful only when it is used in conjunction with other sources of information and policy guidance and in the context of professional experience and judgment. INTERPRETATIONS OF AAMI STANDARDS AND RECOMMENDED PRACTICES Requests for interpretations of AAMI s
22、tandards and recommended practices must be made in writing, to the AAMI Vice President, Standards Policy and Programs. An official interpretation must be approved by letter ballot of the originating committee and subsequently reviewed and approved by the AAMI Standards Board. The interpretation will
23、 become official and representation of the Association only upon exhaustion of any appeals and upon publication of notice of interpretation in the “Standards Monitor“ section of the AAMI News. The Association for the Advancement of Medical Instrumentation disclaims responsibility for any characteriz
24、ation or explanation of a standard or recommended practice which has not been developed and communicated in accordance with this procedure and which is not published, by appropriate notice, as an official interpretation in the AAMI News. American National Standard ANSI/AAMI/ISO 22442-1:2007 Medical
25、devices utilizing animal tissues and their derivativesPart 1: Application of risk management Approved 24 January 2008 by Association for the Advancement of Medical Instrumentation Approved 25 January 2008 by American National Standards Institute, Inc. Abstract: Applies to medical devices other than
26、in-vitro diagnostic medical devices manufactured utilizing materials of animal origin, which are non-viable or have been rendered non-viable. Specifies, in conjunction with ANSI/AAMI/ISO 14971, a procedure to identify the hazards and hazardous situations associated with such devices, to estimate and
27、 evaluate the resulting risks, to control these risks, and to monitor the effectiveness of that control. Outlines the decision process for the residual risk acceptability, taking into account the balance of residual risk, as defined in ANSI/AAMI/ISO 14971, and expected medical benefit as compared to
28、 available alternatives. This part of ANSI/AAMI/ISO 22442 is intended to provide requirements and guidance on risk management related to the hazards typical of medical devices manufactured utilizing animal tissues or derivatives. Keywords: animal tissue, non-viable, transmissible agents, risk manage
29、ment AAMI Standard This Association for the Advancement of Medical Instrumentation (AAMI) standard implies a consensus of those substantially concerned with its scope and provisions. The existence of an AAMI standard does not in any respect preclude anyone, whether they have approved the standard or
30、 not, from manufacturing, marketing, purchasing, or using products, processes, or procedures not conforming to the standard. AAMI standards are subject to periodic review, and users are cautioned to obtain the latest editions. CAUTION NOTICE: This AAMI standard may be revised or withdrawn at any tim
31、e. AAMI procedures require that action be taken to reaffirm, revise, or withdraw this standard no later than five years from the date of publication. Interested parties may obtain current information on all AAMI standards by calling or writing AAMI. All AAMI standards, recommended practices, technic
32、al information reports, and other types of technical documents developed by AAMI are voluntary, and their application is solely within the discretion and professional judgment of the user of the document. Occasionally, voluntary technical documents are adopted by government regulatory agencies or pr
33、ocurement authorities, in which case the adopting agency is responsible for enforcement of its rules and regulations. Published by Association for the Advancement of Medical Instrumentation 1110 N. Glebe Road, Suite 220 Arlington, VA 22201-4795 www.aami.org 2007 by the Association for the Advancemen
34、t of Medical Instrumentation All Rights Reserved This publication is subject to copyright claims of ISO, ANSI, and AAMI. No part of this publication may be reproduced or distributed in any form, including an electronic retrieval system, without the prior written permission of AAMI. All requests pert
35、aining to this draft should be submitted to AAMI. It is illegal under federal law (17 U.S.C. 101, et seq.) to make copies of all or any part of this document (whether internally or externally) without the prior written permission of the Association for the Advancement of Medical Instrumentation. Vio
36、lators risk legal action, including civil and criminal penalties, and damages of $100,000 per offense. For permission regarding the use of all or any part of this document, contact AAMI, 1110 N. Glebe Road, Suite 220, Arlington, VA 22201-4795. Phone: (703) 525-4890; Fax: (703) 525-1067. Printed in t
37、he United States of America Contents Page Glossary of equivalent standards. iv Committee representation vi Background of AAMI Adoption of ISO 22442-1:2007 . vii Foreword ix Introduction x 1 Scope 1 2 Normative references . 2 3 Terms and definitions. 2 4 Risk management process 3 4.1 General. 3 4.2 R
38、isk analysis . 4 4.3 Risk evaluation 5 4.4 Risk control . 5 4.5 Evaluation of overall residual risk acceptability 7 4.6 Production and post-production information system. 7 Annex A (informative) Guidance on the application of this part of ISO 22442. 8 Annex B (informative) Graphical representation o
39、f part of the risk management process for medical devices utilizing animal material. 9 Annex C (normative) Special requirements for some animal materials considering the risk management for TSE agents .11 Annex D (informative) Information relevant to the management of TSE risk.17 Bibliography.25 iv
40、2007 Association for the Advancement of Medical Instrumentation ? ANSI/AAMI/ISO 22442-1:2007 Glossary of equivalent standards International Standards adopted in the United States may include normative references to other International Standards. For each International Standard that has been adopted
41、by AAMI (and ANSI), the table below gives the corresponding U.S. designation and level of equivalency to the International Standard. NOTE: Documents are sorted by international designation. Other normatively referenced International Standards may be under consideration for U.S. adoption by AAMI; the
42、refore, this list should not be considered exhaustive. International designation U.S. designation Equivalency IEC 60601-1:2005 ANSI/AAMI ES60601-1:2005 Major technical variations IEC 60601-1-2:2007 ANSI/AAMI/IEC 60601-1-2:2007 Identical IEC 60601-2-2:2006 ANSI/AAMI/IEC 60601-2-2:2006 Identical IEC 6
43、0601-2-4:2002 ANSI/AAMI DF80:2003 Major technical variations IEC 60601-2-19:1990 and Amendment 1:1996 ANSI/AAMI II36:2004 Major technical variations IEC 60601-2-20:1990 and Amendment 1:1996 ANSI/AAMI II51:2004 Major technical variations IEC 60601-2-21:1994 and Amendment 1:1996 ANSI/AAMI/IEC 60601-2-
44、21 and Amendment 1:2000 (consolidated texts) Identical IEC 60601-2-24:1998 ANSI/AAMI ID26:2004 Major technical variations IEC 60601-2-47:2001 ANSI/AAMI EC38:2007 Major technical variations IEC 60601-2-50:2001 ANSI/AAMI/IEC 60601-2-50:2006 Identical IEC/TR 60878:2003 ANSI/AAMI/IEC TIR60878:2003 Ident
45、ical IEC/TR 62296:2003 ANSI/AAMI/IEC TIR62296:2003 Identical IEC 62304:2006 ANSI/AAMI/IEC 62304:2006 Identical IEC/TR 62348:2006 ANSI/AAMI/IEC TIR62348:2006 Identical ISO 5840:2005 ANSI/AAMI/ISO 5840:2005 Identical ISO 7198:1998 ANSI/AAMI/ISO 7198:1998/2001/(R)2004 Identical ISO 7199:1996 ANSI/AAMI/
46、ISO 7199:1996/(R)2002 Identical ISO 8637:2004 ANSI/AAMI RD16:2007 Major technical variations ISO 8638:2004 ANSI/AAMI RD17:2007 Major technical variations ISO 10993-1:2003 ANSI/AAMI/ISO 10993-1:2003 Identical ISO 10993-2:2006 ANSI/AAMI/ISO 10993-2:2006 Identical ISO 10993-3:2003 ANSI/AAMI/ISO 10993-3
47、:2003 Identical ISO 10993-4:2002 and Amendment 1:2006 ANSI/AAMI/ISO 10993-4:2002 and Amendment 1:2006 Identical ISO 10993-5:1999 ANSI/AAMI/ISO 10993-5:1999 Identical ISO 10993-6:2007 ANSI/AAMI/ISO 10993-6:2007 Identical ISO 10993-7:1995 ANSI/AAMI/ISO 10993-7:1995/(R)2001 Identical ISO 10993-9:1999 A
48、NSI/AAMI/ISO 10993-9:1999/(R)2005 Identical ISO 10993-10:2002 and Amendment 1:2006 ANSI/AAMI BE78:2002 ANSI/AAMI BE78:2002/A1:2006 Minor technical variations Identical ISO 10993-11:2006 ANSI/AAMI/ISO 10993-11:2006 Identical ISO 10993-12:2007 ANSI/AAMI/ISO 10993-12:2007 Identical ISO 10993-13:1998 AN
49、SI/AAMI/ISO 10993-13:1999/(R)2004 Identical ISO 10993-14:2001 ANSI/AAMI/ISO 10993-14:2001/(R)2006 Identical ISO 10993-15:2000 ANSI/AAMI/ISO 10993-15:2000/(R)2006 Identical ISO 10993-16:1997 ANSI/AAMI/ISO 10993-16:1997/(R)2003 Identical International designation U.S. designation Equivalency ISO 10993-17:2002 ANSI/AAMI/ISO 10993-17:2002 Identical ISO 10993-18:2005 ANSI/AAMI BE83:2006 Major technical variations ISO/TS 10993-19:2006 ANSI/AAMI/ISO TIR10993-19:2006 Identical ISO/TS 10993-20:2006 ANSI/AAMI/ISO TIR10993-20:2006 Identical ISO 11135-1:2007 ANSI/AAMI/ISO 11135-1:2007 Ident