《BS-ISO-10651-5-2006.pdf》由会员分享,可在线阅读,更多相关《BS-ISO-10651-5-2006.pdf(54页珍藏版)》请在三一文库上搜索。
1、BRITISH STANDARD BS ISO 10651-5:2006 Lung ventilators for medical use Particular requirements for basic safety and essential performance Part 5: Gas-powered emergency resuscitators ICS 11.040.10 ? Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Lice
2、nsee=Boeing Co/5910770001 Not for Resale, 08/14/2008 22:10:40 MDTNo reproduction or networking permitted without license from IHS -,-,- BS ISO 10651-5:2006 This British Standard was published under the authority of the Standards Policy and Strategy Committee on 30 May 2008 BSI 2008 ISBN 978 0 580 53
3、750 9 National foreword This British Standard is the UK implementation of ISO 10651-5:2006. The UK participation in its preparation was entrusted by Technical Committee CH/121, Anaesthetic and respiratory equipment, to Subcommittee CH/121/5, Lung ventilators, tracheal tubes and related equipment. A
4、list of organizations represented on this committee can be obtained on request to its secretary. This publication does not purport to include all the necessary provisions of a contract. Users are responsible for its correct application. Compliance with a British Standard cannot confer immunity from
5、legal obligations. Amendments/corrigenda issued since publication DateComments Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Licensee=Boeing Co/5910770001 Not for Resale, 08/14/2008 22:10:40 MDTNo reproduction or networking permitted without licen
6、se from IHS -,-,- Reference number ISO 10651-5:2006(E) INTERNATIONAL STANDARD ISO 10651-5 First edition 2006-02-01 Lung ventilators for medical use Particular requirements for basic safety and essential performance Part 5: Gas-powered emergency resuscitators Ventilateurs pulmonaires usage mdical Exi
7、gences particulires pour la scurit de base et les performances essentielles Partie 5: Appareils de ranimation durgence aliments par gaz BS ISO 10651-5:2006 Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Licensee=Boeing Co/5910770001 Not for Resale,
8、 08/14/2008 22:10:40 MDTNo reproduction or networking permitted without license from IHS -,-,- ii Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Licensee=Boeing Co/5910770001 Not for Resale, 08/14/2008 22:10:40 MDTNo reproduction or networking perm
9、itted without license from IHS -,-,- iii Contents Page Foreword iv Introduction v 1 * Scope 1 2 Normative references. 1 3 Terms and definitions. 2 4 General requirements. 4 4.1 General. 4 4.2 Other test methods. 5 4.3 Acceptance criteria. 5 5 Constructional requirements. 6 5.1 General. 6 5.2 Connect
10、ors 7 6 Operational requirements 9 6.1 General. 9 6.2 Resistance to environmental influences 10 6.3 Resistance to typical use. 10 6.4 Size and mass . 11 6.5 Gas supply. 11 7 Performance 12 7.1 Ventilatory requirements . 12 7.2 * Ventilation performance . 14 8 Identification, marking and documents 17
11、 8.1 General. 17 8.2 Marking 18 8.3 Information provided by the manufacturer for operation and maintenance 19 Annex A (informative) Rationale. 23 Annex B (normative) Test methods 28 Annex C (informative) Relationship between this part of ISO 10651 and cardiopulmonary resuscitation guidelines. 40 Ann
12、ex D (informative) Clauses of this part of ISO 10651 addressing the essential principles 41 Annex E (informative) Terminology Index of defined terms 43 Bibliography. 45 BS ISO 10651-5:2006 Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Licensee=Boe
13、ing Co/5910770001 Not for Resale, 08/14/2008 22:10:40 MDTNo reproduction or networking permitted without license from IHS -,-,- iv Foreword ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing Inter
14、national Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison wit
15、h ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization. International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2. The main task of technica
16、l committees is to prepare International Standards. Draft International Standards adopted by the technical committees are circulated to the member bodies for voting. Publication as an International Standard requires approval by at least 75 % of the member bodies casting a vote. Attention is drawn to
17、 the possibility that some of the elements of this document may be the subject of patent rights. ISO shall not be held responsible for identifying any or all such patent rights. ISO 10651-5 was prepared by Technical Committee ISO/TC 121, Anaesthetic and respiratory equipment, Subcommittee SC 3, Lung
18、 ventilators and related equipment. This first edition of ISO 10651-5, together with ISO 10651-4:2002, cancels and replaces ISO 8382:1988, which has been technically revised. ISO 10651 consists of the following parts, under the general title Lung ventilators for medical use Particular requirements f
19、or basic safety and essential performance: Part 2: Home care ventilators for ventilator-dependent patients Part 3: Particular requirements for emergency and transport ventilators Part 4: Particular requirements for operator-powered resuscitators Part 5: Gas-powered emergency resuscitators Part 6: Ho
20、me-care ventilatory support devices NOTE ISO 10651-1:1993, Lung ventilators for medical use Part 1: Requirements, was withdrawn in 2001 and has been revised as IEC 60601-2-12:2001, Medical electrical equipment Part 2-12: Particular requirements for the safety of lung ventilators Critical care ventil
21、ators. BS ISO 10651-5:2006 Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Licensee=Boeing Co/5910770001 Not for Resale, 08/14/2008 22:10:40 MDTNo reproduction or networking permitted without license from IHS -,-,- v Introduction For victims whose l
22、ives are at risk from respiratory failure, in particular during cardiac arrest, resuscitation councils and associations teach that the best ultimate outcome will be achieved if there is a continuous chain of care starting with earliest possible bystander cardiopulmonary resuscitation and continuing
23、until the victim can be put under professional medical care. In order to improve the care possible at the early stages of this chain, authorities and organizations are training non-specialized personnel in key situations, such as where people congregate or where there are increased risks, so that th
24、ey can be available to provide a higher level of care with a minimum of delay. There is a growing realization that the effectiveness of such intervention can be greatly enhanced by the use of certain basic equipment, such as that which provides ventilation whilst avoiding mouth-to-mouth contact. Sim
25、ple, gas-powered emergency resuscitators can deliver controlled ventilation for this purpose and this document specifies the criteria they are required to satisfy. In this part of ISO 10651, the following symbols and notations are used: requirements, compliance with which can be tested, and definiti
26、ons: roman type; notes, explanations, advice, introductions, general statements and references: smaller roman type; test specifications: italic type; terms defined in ISO 4135:2001, IEC 60601-1:1988 or in this part of ISO 10651: bold type. Throughout this part of ISO 10651, text for which a rational
27、e is provided in Annex A is indicated by an asterisk (*). BS ISO 10651-5:2006 Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Licensee=Boeing Co/5910770001 Not for Resale, 08/14/2008 22:10:40 MDTNo reproduction or networking permitted without licens
28、e from IHS -,-,- blank Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Licensee=Boeing Co/5910770001 Not for Resale, 08/14/2008 22:10:40 MDTNo reproduction or networking permitted without license from IHS -,-,- 1 Lung ventilators for medical use Par
29、ticular requirements for basic safety and essential performance Part 5: Gas-powered emergency resuscitators 1 * Scope This part of ISO 10651 specifies the basic safety and essential performance requirements for gas-powered emergency resuscitators (3.10) intended for use with humans by first responde
30、rs. This equipment is intended for emergency field use and is intended to be continuously operator attended in normal use. This part of ISO 10651 also specifies the requirements for resuscitator sets (3.22). This part of ISO 10651 is not applicable to electrically-powered resuscitators. NOTE ISO 106
31、51-3 covers emergency and transport ventilators. 2 Normative references The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including an
32、y amendments) applies. ISO 31 (all parts), Quantities and units ISO 32, Gas cylinders for medical use Marking for identification of content ISO 1000, SI units and recommendations for the use of their multiples and of certain other units ISO 4135:2001, Anaesthetic and respiratory equipment Vocabulary
33、 ISO 5356-1, Anaesthetic and respiratory equipment Conical connectors Part 1: Cones and sockets ISO 5356-2, Anaesthetic and respiratory equipment Conical connectors Part 2: Screw-threaded weight- bearing connectors ISO 5359, Low-pressure hose assemblies for use with medical gases ISO 5367, Breathing
34、 tubes intended for use with anaesthetic apparatus and ventilators ISO 9170-1, Terminal units for medical gas pipeline systems Part 1: Terminal units for use with compressed medical gases and vacuum ISO 10297, Gas cylinders Refillable gas cylinder valves Specification and type testing ISO 10524-1, P
35、ressure regulators for use with medical gases Part 1: Pressure regulators and pressure regulators with flow-metering devices BS ISO 10651-5:2006 Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Licensee=Boeing Co/5910770001 Not for Resale, 08/14/2008
36、 22:10:40 MDTNo reproduction or networking permitted without license from IHS -,-,- 2 ISO 10524-3, Pressure regulators for use with medical gases Part 3: Pressure regulators integrated with cylinder valves ISO 11607, Packaging for terminally sterilized medical devices ISO 14971, Medical devices Appl
37、ication of risk management to medical devices and Amendment 1:2003 ISO 15223:2000, Medical devices Symbols to be used with medical device labels, labelling and information to be supplied and Amendment 1:2002 and Amendment 2:2004 ISO 17664, Sterilization of medical devices Information to be provided
38、by the manufacturer for the processing of resterilizable medical devices ISO 23328-1:2003, Breathing system filters for anaesthetic and respiratory use Part 1: Salt test method to assess filtration performance ISO 23328-2:2002, Breathing system filters for anaesthetic and respiratory use Part 2: Non
39、-filtration aspects IEC 60529:2001, Degrees of protection provided by enclosures (IP code) IEC 60601-1:1988, Medical electrical equipment Part 1: General requirements for safety and Amendment 1:1991 and Amendment 2:1995 3 Terms and definitions For the purposes of this document, the terms and definit
40、ions given in ISO 4135:2001, IEC 60601-1:1988 and the following apply. For convenience, the sources of all defined terms used in this document are given in Annex E. 3.1 accompanying documents documents accompanying resuscitator or resuscitator sets and containing all important information for the us
41、er, operator, installer or assembler of the resuscitator, particularly regarding safety NOTE Adapted from IEC 60601-1:1988, definition 2.1.4. 3.2 automatic pressure-cycled resuscitator resuscitator in which the cycling from the inspiratory phase to the expiratory phase occurs after attaining a press
42、ure determined by the control setting 3.3 automatic time-cycled resuscitator resuscitator in which the cycling between the inspiratory phase and expiratory phase is controlled automatically at time intervals determined by the control setting 3.4 automatic volume-cycled resuscitator resuscitator in w
43、hich the cycling from the inspiratory phase to the expiratory phase occurs after the delivery of a delivered volume determined by the control setting 3.5 cardiopulmonary resuscitation combination of rescue breathing and chest compressions delivered to victims thought to be in cardiac arrest AHA Guid
44、elines for Cardiopulmonary Resuscitation and Emergency Care BS ISO 10651-5:2006 Copyright British Standards Institution Provided by IHS under license with BSI - Uncontrolled Copy Licensee=Boeing Co/5910770001 Not for Resale, 08/14/2008 22:10:40 MDTNo reproduction or networking permitted without lice
45、nse from IHS -,-,- 3 3.6 clearly legible capable of being read by the operator or other relevant person with normal vision NOTE See also 8.2.1. 3.7 * delivered volume Vdel volume of gas delivered through the patient connection port during an inspiratory phase 3.8 demand valve part of the resuscitato
46、r that delivers a flow of gas related to a reduction of pressure generated by the patient at the patient connection port 3.9 first responder individual who has been trained to provide primary response to a respiratory emergency EXAMPLE 1 Fire fighter. EXAMPLE 2 Emergency medical technician. 3.10 gas
47、-powered emergency resuscitator resuscitator portable equipment, powered by compressed gas, intended for immediate use to provide lung ventilation in the resuscitation of individuals who have sudden breathing difficulties 3.11 inadvertent positive end-expiratory pressure inadvertent PEEP unintended positive pressure at the patient connection port at the end of the expiratory phase 3.12 intermediate hose hose that conducts gas between parts of the resuscitator 3.13 manually-cycled resuscitator resuscitator in which the inspiratory phase and expiratory phase are controlled by a repe